From time to time TTi may make updates to
the firmware of the digital bus interface available.
This firmware resides in Flash RAM and can
be updated using the USB interface.
How to check the current firmware version
This page provides information and downloadable files relevant to
the New PL-P and PLH-P Series of laboratory dc power supplies.
The interface firmware level is displayed very briefly when the
power supply is switched on. It is the second number that
is briefly displayed and is of the form IF X.XX (e.g. IF 1.01).
For full compliance with LXI 1.2, and to
achieve other improvements in interface performance, earlier
versions of the interface firmware should be updated
accordingly. The latest firmware version is IF 2.13,
and for those using the LAN interface updating to this version
is recommended. Important Note: If updating from
an earlier version, an updated version of the User Manual should
also be downloaded in order that correct operation of the remote
interfaces is maintained.
Go Here to download the
latest User Manual.
Important change introduced from IF 1.04 onwards
It is an LXI Compliance requirement that front panel indication
is given to the user if no Physical LAN connection is found by
the instrument when it is switched on. The
implementation of this was introduced in IF 1.04, and is
described in Issue 2 of the Instruction Manual under "Switching
On and Power-On Conditions".
Go Here to download the
latest User Manual.
|