WAVEFORMS
Sine
Frequency Range: 1µHz to 50MHz (TG2511
- 1µHz to 25MHz)
Frequency Resolution: 1µHz, 14 digits
Output Level: 10mVp-p to 10Vp-p into 50Ohms
Amplitude Flatness (Relative to 1kHz): <100kHz 0.1dB, <5MHz
0.15dB, <20MHz 0.3dB, <50MHz 0.5dB
Harmonic Distortion: <1 Vp-p
≥ 1Vp-p
DC to 20kHz
-65dBc
-65dBc
20kHz to 100kHz
-60dBc
-60dBc
100kHz to 1MHz
-45dBc
-45dBc
1MHz to 25MHz
-40dBc
-35dBc
25MHz to 50MHz
-40dBc
-28dBc
Non-Harmonic Spurii: <–60dBc to 1MHz, <–60dBc + 6dB/octave 1MHz
to 50MHz
Phase Noise (10kHz offset): -115dBc/Hz, typical
Square
Frequency Range: 1µHz to 50MHz (TG2511
- 1µHz to 25MHz)
Frequency Resolution: 1µHz, 14 digits
Output Level: 10mVp-p to 10Vp-p into 50Ohms
Rise and Fall Times: <8ns (TG2511 - <13ns)
Overshoot: <5%
Variable Duty Cycle: 20% to 80% to 10MHz, 0.1% resolution, 40%
to 60% to 25 MHz, 0.1% resolution
50% (fixed) above 25MHz
Asymmetry (@ 50% duty): 1% of period + 5ns
Jitter (RMS) 0.5ns + 100 ppm of period
Ramp and Triangle
Frequency Range: 1µHz to 500kHz (TG2511
- 1µHz to 250kHz)
Frequency Resolution: 1µHz, 12 digits
Output Level: 10mVp-p to 10Vp-p into 50Ohms
Linearity Error: <0.1% to 30 kHz
Variable Symmetry: 0.0 % to 100.0 %, 0.1% resolution
Single key operation of 50% (Triangle)
Note the triangle and sawtooth waveforms are also available from
the arbitrary waveform menu enabling repetition rates of up to
10MHz/6MHz. Waveform quality will deteriorate at higher
frequencies however.
Pulse
Frequency Range: 500 µHz to 12.5MHz (TG2511
- 500 µHz to 6.25MHz)
Frequency Resolution: 1µHz , 14 digits
Output Level: 10mVp-p to 10Vp-p into 50Ohms
Overshoot: <5%
Jitter: 300ps + 0.01% of period
Rise/Fall Times: Rise and Fall times can be independently varied
or can be varied together simultaneously.
Range: <8ns to 40µs (TG2511 - <13ns to
40µs)
Resolution: 0.1ns for rise/fall time ≤100ns; 1ns for
rise/fall >100ns and ≤2µs; 10ns for rise/fall >2µs and ≤40µs )
Width Range: 20ns to 2000s (20ns minimum for period ≤40s) (200
ns minimum for period >40s and ≤400s)
(2µs minimum for period >400s)
Width Resolution: 10ns (for period ≤40s),100ns (for period >40s
and ≤400s),1µs (for period >400s)
Delay Range: 0ns to 2000s
Delay Resolution: 10ns (for period ≤40s),100ns (for period >40s
and ≤400s),1µs (for period >400s)
Arbitrary
In built arbitrary waveforms (Sinc,
Exponential Rise, Logarithmic Rise, DC, Positive and Negative
Ramps and Square). Additional waveforms are supplied on disc
(Cardiac, Gaussian, Exponential Fall, Logarithmic Fall).
Up to 4 additional or user defined waveforms may be stored in
non-volatile memory. Waveforms can be defined by downloading of
waveform data via USB memory stick, remote interfaces, or
editing via the instrument’s front panel.
Waveform Size: 2 points to 131072 points (128k).
Waveform Memory Size: Up to 4 waveforms of up to 64k points, or
2 waveforms of up to 128k points, (or 2 of 64k points plus 1 of
128k points).
External Waveform Storage: up to 1,000 waveforms per USB memory
stick
Vertical Resolution: 14 bits
Frequency Range: 1µHz to 10MHz (TG2511
- 1µHz to 6MHz)
Frequency Resolution: 1µHz , 14 digits
Output Level: 10mVp-p to 10Vpp into 50Ohms
Sampling rate: 125MS/s
Output Filter: Selects between 50MHz Elliptic or 20MHz Bessel
filter depending on the waveform.
Arbitrary Waveform Creation and Editing
Both generators are supplied with Waveform
Manager Plus. This Windows based software provides a
sophisticated tool set for the creation, editing and management
of arbitrary waveforms. The waveforms can be transferred
to the generator either using a USB memory stick, or by the
digital interfaces.
Noise
Gaussian White Noise: Noise can be added to
any carrier waveform (except pulse, square and noise itself)
note however that noise can be added to the square wave
available in the arbitrary menu. The amount of noise added can
be specified as 0% to 50% of the amplitude of the carrier
waveform. Noise can also be used as modulating waveform.
Bandwidth (-3dB): 20MHz typical.
Noise crest factor (Vp/Vrms): 5.27
Output Level: 10mVp-p to 10Vpp into 50Ohms
Internal Frequency Reference
Oscillator Ageing Rate: 1ppm first year
Temperature Stability: <1ppm over the specified temperature
range
MODULATION
AM
Carrier Waveforms: Sine, Square, Ramp, Arb
Modulation Source: Internal/External
Internal Modulating Waveforms: Sine, Square, Up Ramp, Down Ramp,
Triangle, Noise, DC, Sinc, Exponential Rise, Logarithmic Rise
and User Defined Arbs
Internal Modulating Frequency: 1µHz to 20kHz, 1µHz resolution
Amplitude Depth: 0.0% to 120.0%, 0.1% resolution
FM
Carrier Waveforms: Sine, Square, Ramp, Arb
Modulation Source: Internal/External
Internal Modulating Waveforms: Sine, Square, Up Ramp, Down Ramp,
Triangle, Noise, DC, Sinc, Exponential Rise, Logarithmic Rise
and User Defined Arbs
Internal Modulating Frequency: 1µHz to 20kHz, 1µHz resolution
Frequency Deviation: DC to Fmax/2, 1µHz resolution
PM
Carrier Waveforms: Sine, Square, Ramp, Arb
Modulation Source: Internal/External
Internal Modulating Waveforms: Sine, Square, Up Ramp, Down Ramp,
Triangle, Noise, DC, Sinc, Exponential Rise, Logarithmic Rise
and User Defined Arbs
Internal Modulating Frequency: 1µHz to 20kHz, 1µHz resolution
Phase Deviation: -360.0 to +360.0 degrees, 0.1 degree resolution
PWM
Carrier Waveforms: Pulse
Modulation Source: Internal/External
Internal Modulating Waveforms: Sine, Square, Up Ramp, Down Ramp,
Triangle, Noise, DC, Sinc, Exponential Rise, Logarithmic Rise
and User Defined Arbs
Internal Modulating Frequency: 1µHz to 20kHz, 1µHz resolution
Pulse Width Deviation: 0% to 100% of pulse width, resolution
same as of pulse width
FSK
Carrier Waveforms: Sine, Square, Ramp, Arb
Source: Internal/External (via TRIG IN)
Internal Modulation: 50% duty cycle square (2mHz to 100kHz)
Triggered Burst
Each active edge of the trigger signal will produce one burst of
the waveform.
Carrier Waveforms: Sine, Square, Ramp, Arb, Pulse
Maximum Carrier Frequency: 10MHz (finite cycles), 50MHz
(infinite), subject to carrier waveform.
Number of Cycles: 1 to 1,048,575 and infinite.
Trigger Repetition Rate: 2mHz to 1MHz internal dc to 1MHz
external.
Trigger Signal Source: Internal from keyboard or trigger
generator. External from TRIG IN or remote interface.
Trigger Start/Stop Phase: -360.0 to +360.0 degrees, 0.1 degree
resolution, subject to carrier waveform
Gated
Waveform will run while the Gate signal is true and stop while
false.
Carrier Waveforms: Sine, Square, Ramp, Arb, Pulse, Noise
Maximum Carrier Frequency: 10 MHz, subject to carrier waveform
Trigger Repetition Rate: 2mHz to 1MHz internaldc to 1MHz
external.
Gate Signal Source: Internal from keyboard or trigger generator.
External from TRIG IN or remote interface.
Gate Start/Stop Phase: -360.0 to +360.0 degrees, 0.1 degree
resolution, subject to carrier waveform
Sweep
Frequency sweep capability is provided for both standard and
arbitrary waveforms.
Carrier Waveforms: All standard and arbitrary except pulse.
Sweep Mode: Linear or logarithmic, triggered or continuous.
Sweep Direction: Up, down, up/down or down/up.
Sweep Range: From 1µHz to 50MHz, subject to carrier waveform.
Phase continuous. Independent setting of the start and stop
frequency.
Sweep Time: 1ms to 500s (6 digit resolution).
Marker: Variable during sweep.
Sweep Trigger Source: The sweep may be free run or triggered
from the following sources: Internal from keyboard or trigger
generator. Externally from TRIG IN input or remote interface.
Trigger Generator
Internal source 2mHz to 1MHz square wave adjustable in 1us
steps, 9 digit resolution. Available for external use from the
SYNC OUT socket.
OUTPUTS
Main Output
Output Impedance: 50Ohms
Amplitude: 20mV to 20Vp-p open circuit (10mV to 10Vp-p into
50Ohms). Amplitude can be specified open circuit (hi Z) or into
an assumed load of 1Ohm to 10kOhm in Vpk-pk, Vrms or dBm.
Amplitude Accuracy: 2% ±1mV at 1kHz into 50Ohm.OhmW.
DC Offset Accuracy: Typically 3% ±10mV.
Resolution: 3 digits or 1mV for both Amplitude and DC Offset.
Sync Out
Multifunction output user definable or automatically selected to
be any of the following:
Carrier Waveform Sync: Sine/Ramp/Pulse - A square wave with 50%
duty cycle at the waveform frequency.
Square - A square wave with same duty cycle as the main output
at the waveform frequency.
Arbs - A square wave with 50% duty cycle at the waveform
frequency. The sync is a TTL high when the first point of the
waveform is output. Noise - No sync associated with noise.
Modulation Sync: AM/FM/PM/PWM A square wave with 50% duty cycle
referenced to the internal modulation waveform when modulation
source is internal, or a square wave referenced to the carrier
waveform when modulation source is external. No sync is
associated with noise as the modulation source.
FSK - A square wave referenced to the trigger rate. The
sync is a TTL high when hop frequency is the output frequency
and TTL low when carrier frequency is the output frequency for
positive slope and vice versa for negative slope.
Burst Sync: A square wave that is a TTL high when the burst
begins and a TTL low when burst is completed.
Trigger: Selects the current trigger signal. Useful for
synchronizing burst or gated signals.
Sweep Sync: Marker Off - A square wave that is a TTL low from
the midpoint of the sweep and a TTL high from the end of the
sweep. Marker On - A square wave that is a TTL low from the
marker frequency and a TTL high from the end of the sweep.
Output Signal Level: Logic level nominally 3V.
Ref Clock Output
Buffered version of the 10MHz clock currently in use (internal
or external)
Output Level: Nominally 3V logic level from 50Ohm.
INPUTS
Trig In
Frequency Range: DC - 1MHz.
Signal Range: Threshold nominally TTL level; maximum input ±10V.
Minimum Pulse Width: 50ns
Polarity: Selectable as high/rising edge or low/falling edge.
Input Impedance: 10kOhm
External Modulation Input (for AM,
FM, PM, PWM)
Voltage Range: ± 5V full scale
Input Impedance: 5kOhm typical
Bandwidth: DC to 20kHz
Ref Clock Input
Input for an external 10MHz reference clock
Voltage Range: 1Vpp – 5Vpp
Maximum Voltage: +5V
Minimum Voltage: -1V
Phase synchronising two generators
Two generators can be synchronised together
to provide outputs at the same frequency (or harmonics) and with
a phase difference. The amplitude and phase of these outputs can
also be modulated providing the capability to perform QAM and
QPSK respectively. It is also possible to synchronise more than
two generators but results are not guaranteed.
Carrier Waveforms: Sine, Square, Ramp, Pulse, Arb
Phase: Range -360.0 to +360.0 degrees
Resolution 0.1 degree
Accuracy < ±5ns
INTERFACES
Full digital remote control facilities are
available through LAN and USB and optional GPIB interfaces.
LAN Interface - Ethernet 100/10base – T hardware connection. LXI
V1.2, Class C compliant.
USB Interface - Standard USB 2.0 hardware connection.
Implemented as virtual-COM port.
GPIB Interface (option G) - conforming with IEEE-488.2
USB Flash Drive For waveform and setup storage/recall.
GENERAL
Display: Type: Monochrome Graphics Display -
pixel format: 112 X 256 Matrix
Data Entry: Keyboard selection of mode, waveform etc.; value
entry direct by numeric keys or by rotary control.
Stored Settings: Up to 9 complete instrument set-ups may be
stored and recalled from non-volatile memory.
Size: Bench top use: 97mm height; 250mm width; 270mm long,
Rack mounted use: 86.5mm (2U) height;
213.5mm (½-rack) width; 244mm long
Weight: 2.55 kg
Power: 110-240VAC ±10% 50/60Hz; 100-120VAC ±10% 400Hz; 60VA max.
Installation Category II.
Operating Range: +5°C to 40°C, 20-80% RH.
Storage Range: -20°C to + 60°C.
Environmental: Indoor use at altitudes up to 2000m, Pollution
Degree 2.
Safety: Complies with EN61010-1.
EMC: Complies with EN61326
OPTIONS Product with GPIB
interface (option G)
19 inch rack mounting kit. SUPPLIED ITEMS
Each product is supplied with the following:
Printed operating manual in English, French, German, Italian and
Spanish with remote programming instruction in English only. PDF
operating manuals on disk (including remote programming
instruction) in English, French, German, Italian and Spanish.
IEC mains lead appropriate to the geographic market in which it
is sold.
Software on disk including: Waveform Manager for Windows, IVI
driver, LabView driver, CVI driver, USB driver, LXI discovery
tool, TCP example program. Specifications
apply for the temperature range 18°C to 28°C after 30 minutes
warm-up, at maximum output into 50 Ohms unless otherwise stated.
Thurlby Thandar Instruments Ltd. operates a policy of continuous
development and reserves the right to alter specifications
without prior notice. |