Signal and Waveform Generator Products from TTi - Thurlby Thandar Instruments

  TG200 Series - Description and Specifications

TG210 - 2MHz Dial-set Function Generator

  0.02Hz to 2MHz frequency range
  High waveform quality at all frequencies & levels
  20mV to 20V pk-pk from 50 Ohms or 600 Ohms
  Sine, square and triangle waveforms plus dc
  1000:1 frequency change on each range
  Variable symmetry control
 
External sweep input


 

N.B. The TG215 and TG230 models have been
 discontinued and replaced by the TG315 and TG330

 

   Larger Image

Three models meeting a variety of needs

The TG210 offers all of the basic generator functions including sine, square & triangle waveforms, variable DC offset, variable symmetry and external sweep input.

The TG215 adds a digital display of frequency, amplitude and offset for greater precision and convenience.

The TG230 has all the features of the TG215 plus an internal sweep generator with linear or log sweep, variable sweep rate and a sweep output socket. It also offers internal or external amplitude modulation.

Exceptional waveform quality
The TG200 series provides very high waveform quality under all conditions. That means low sinewave distortion, low aberation triangle waves and fast-edged square waves with low overshoot.
Unlike many competitive products good waveform quality is maintained throughout the frequency range, and at low output levels.
All models have bi-directional variable symmetry at constant frequency from 1:9 to 9:1.
Digital display for precision and convenience
The TG215 and TG230 have the convenience of a digital display. Unlike competitive products the display provides a readout not just of frequency, but of amplitude and offset as well.


Wide range level control
The main output has a maximum emf of 20V pk-pk from a 50Ohm or 600Ohm source. An amplitude vernier with a range of 40dB is combined with a -20dB switched attenuator to provide levels down to 20mV pk-pk. Variable DC offset of +/-10V is available via a centre detent control.
An auxiliary output provides a fixed 0V to +5V level suitable for driving both TTL and CMOS loads.

Wide sweep range
Each range can be swept by at least 1000:1 either manually or via the sweep control input. The TG230 incorporates a versatile internal sweep generator capable of providing linear or logarithmic frequency sweeps.
The TG230 provides further versatility by adding AM facilities. The modulation can be external or internal (via a 400Hz internal generator) with modulation levels fully variable between 0% and 100%.
Frequency
Frequency Range <0.02Hz to 2MHz in 7 overlapping decade ranges with fine adjustment by a vernier.
Vernier Range: 1000:1 on each range.
Vernier Accuracy: Typically +/-5% of full scale for TG210; see Meter Accuracy for TG215 & TG230.
Sweep Modes
Internal (TG230 only)
Sweep Range: 1000:1 within each range.
Sweep Rate: Adjustable, typically 20ms to 20 secs.
Sweep Mode: Linear or logarithmic. Sweep start and stop frequencies displayed at the press of a button.
External
Input Impedance: 10kW
Input Sensitivity: 0 to 3V for 1000:1 sweep
Max. Input Voltage: +/-10V
Sweep Linearity: Better than 1%
Max. Slew Rate: 0.1V/us
Amplitude Modulation (TG230 only)
Depth: Variable 0 to 100%
Frequency: 400Hz (internal). DC to 100kHz (external).
External Sensitivity: Approximately 2V peak-to-peak for 50% modulation.
Meter Accuracy (TG215/230 only)
Frequency: +/- 1 digit on 2k to 2M ranges; better than 1% of full scale on 2 to 200 ranges.
Amplitude: Typically 5% of full scale
DC Offset: Typically 2%
Resolution: 0.05% maximum on all ranges (1 in 2000)
Update Rate: Varies with range; at least 2 updates per second
Operating Modes
Specifications apply for the top decade of each frequency range and output 10V peak-to-peak into 50W termination. 
Sine Distortion: Less than 0.5% on 200, 2k and 20k ranges; less than 1% on 2, 20 and 200k ranges; all harmonics >25dB below fundamental on 2M range.
Amplitude Flatness: +/-0.2dB to 200kHz; +/-1dB to 2MHz.
Triangle Linearity: Better than 99% to 200kHz
Rise and Fall Times: <100ns
Mark - Space Ratio: 1:1 +/- 1% to 100kHz
DC Range: +/-10V from 50W
Symmetry Range: Variable typically between 1:9 to 9:1 (on top decade of each range), frequency divided by 10.
Outputs
50W and 600W
Two switch-selectable ranges with 40dB vernier control within each range. (No 600W output on TG230)
0dB Range: 0.2V to 20V peak-to-peak (0.1V to 10V into 50W).
-20dB Range: 20mV to 2V peak-to-peak (10mV to 1V into 50W)
DC Offset Control Range: +/-10V from 50W. DC offset plus signal peak limited to +/-10V (+/-5V into 50W). DC offset plus waveform attenuated proportionally in -20dB position.
TTL
Fixed 0 to +5V output capable of driving CMOS and 2 standard TTL loads.
Sweep Out
6V ramp from 600W.
General
Input Voltage: 110/120 volts AC nominal 50/60Hz or 220/240 volts AC nominal 50/60Hz by rear panel adjustment. 25VA max.
Power Consumption: 25VA max.
Operating Temperature Range: +5oC to +40oC, 20% to 80% RH.
Storage Temperature Range: -10oC to +65oC
Size: 260(W) x 88(H) x 235(D)mm (10.2 x 3.4 x 9.2"), excluding handle and feet.
Weight: 1.9kg (4.2lb)
 Thurlby Thandar Instruments Ltd. operates a policy of continuous development and reserves the right to alter specifications 
  without prior notice.

Thurlby Thandar Instruments Limited
Glebe Road, Huntingdon, Cambridgeshire PE29 7DR United Kingdom
Tel: +44 1480 412451  Fax: +44 1480 450409  Web: www.tti-test.com  Email: sales@tti-test.com